| Vacuum Test Packs
Raven Vacuum Test Pack (a Bowie-Dick type test) is a disposable pack used to evaluate the efficiency of the vacuum system of a pre-vacuum steriliser. Performed daily, this test is a quick indication of steriliser malfunction. This test is not a sterility test, but it is an integral part of a comprehensive quality assurance program. This test verifies that the steriliser's vacuum phase is removing a sufficient amount of air prior to the introduction of steam into the chamber. It also tests for air leaks into the chamber during the sterilisation cycle. The Raven logo on the outside of the pack is printed with a chemical indicator. When exposed to steam, it will change from blue to brown/black. This is to aid the technician in differentiating exposed from unexposed packs. If the Raven logo is not blue prior to use, do not use pack. This back is not intended to evaluate if sterilisation conditions have been achieved. Raven offers other indicators specifically for this purpose. Raven Vacuum Test Pack meets with ANSI/AAMI ST46-1993 standards as published in AAMI Standards and Recommended Practices, 1995. Procedure for use
Interpreting the testA FAIL is indicate by a lighter colour in the centre of the pattern than around the edges - failure of the pattern to change completely to black. A PASS is obvious by a uniform black colour over the entire pattern of the indicator sheet. PackagingRaven Vacuum Test Pack is packaged 80 tests per case. |